A Elbow body
B Elbow ring
C Headgear
D Frame (SLM)
E Cushion
F Elbow clip
1 Vent
English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient re-use of the AirFit N30i nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning
instructions. The processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
AirFit N30i mask component
1
Cleaning - Alconox
Validated number of cycles
Thermal Disinfection - Manual
194-199°F (90-93°C) for 1 to 10 minutes
Cushion
30
Frame (SLM)
2
30
Elbow body / elbow clip / elbow ring
30
Headgear
30
1
This mask may not be available in all regions. For full details regarding the correct use of this mask, please refer to the specific User Guide. For a list of available replacement parts for each mask system, check the Product Guide on ResMed.com.
2
The frame of the Sleep Lab Mask (SLM) does not contain sleeves.
A
B
D
1
1
E
C
F
DRAFT 20180612 608485 AirFit N30i Disinfection Guide AMER Eng Ref.indd 1 13/06/2018 10:48:12 AM

Other manuals for ResMed AirFit N30i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirFit N30i and is the answer not in the manual?

ResMed AirFit N30i Specifications

General IconGeneral
Cushion MaterialSilicone
ManufacturerResMed
Mask TypeNasal Cradle
Headgear IncludedYes
Size OptionsSmall, Medium, Wide
Compatible DevicesCompatible with most CPAP machines
WeightApprox. 2.08 oz

Related product manuals