EasyManuals Logo

ResMed AirSense 11 Disinfection Guide

ResMed AirSense 11
6 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
1
Disinfection Guide
Preparing the device for use between patients
When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health
care provider, the outlet connector, air tubing , cleanable humidifier tub and device enclosure should be
reprocessed between each patient.
Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the
outlet connector, air tubing, cleanable humidifier tub and device enclosure. The person executing
reprocessing activities is responsible for ensuring that reprocessing is completed in line with ResMed's
validated procedures. Components not identified in the reprocessing instructions do not require
reprocessing or are intended for single patient use.
Note: The standard HumidAir 11 tub cannot be reprocessed.
WARNING
Always follow cleaning and disinfection instructions. Some cleaning products may damage
the components and their function or leave harmful residual vapors.
Any deviations from the procedures or claimed maximum number of cycles in this guide can
have an adverse effect on the components and consequently the safety or the quality of
therapy.
When using detergents, disinfectants or equipment, always follow the instructions provided
by the manufacturer of those products. In the event of conflict, this guide takes precedence.
Always follow safe operating practices, including the use of appropriate Personal Protective
Equipment (PPE), as required. Refer to the instructions provided by the manufacturer of those
products for more details.
Beware of electrocution:
Do not immerse the device, AC Adaptor or power cord in water.
Do not connect to power while the device is wet. Make sure that all parts are dry before plugging
it in.

Other manuals for ResMed AirSense 11

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirSense 11 and is the answer not in the manual?

ResMed AirSense 11 Specifications

General IconGeneral
Device TypeCPAP Machine
ModesCPAP, AutoSet, AutoSet for Her
HumidificationIntegrated HumidAir humidifier
DisplayColor LCD touch screen
Pressure Range4-20 cm H2O
Water Chamber Capacity380 mL
Altitude AdjustmentAutomatic
ConnectivityBluetooth, Cellular
Weight1.13 kg (2.49 lbs) (device with water tub)
Heated Tubing OptionClimateLineAir
Data StorageSD Card
Power SupplyAC 100-240V, 50-60Hz
Operating Conditions5°C to 35°C (41°F to 95°F)
CompatibilityResMed masks
Ramp Time0-45 minutes (AutoRamp available)
Dimensions259 x 129 x 150 mm (10.2 x 5.1 x 5.9 inches)

Related product manuals