English 3
Decontaminating
Before the disinfection process, each component must be cleaned and rinsed so no visible contamination is
present.
1. Clean all components with a soft bristle brush for one minute while soaking them in the detergent
solution. Pay particular attention to all crevices and cavities.
2. Thoroughly rinse each component according to the manufacturer’s instructions.
ResMed has tested the following detergents according to the manufacturers instructions:
Detergent Water temperature
Alconox™ Tergazyme
(diluted at 1%)
Hot water (approx 60°C)
Warm water (approx 50°C)
Aniosyme™ DDI
(diluted at 0.5%)
Room temperature water (approx 20°C)
Neodisher MediZym™
(diluted at 2.0%)
Warm water (approx 45°C)
Disinfecting
In the procedures below, only one disinfection or sterilization procedure needs to be performed:
High level thermal disinfection
High level chemical disinfection
Sterilization.
High level thermal disinfection
1. Soak the disassembled components in a hot water bath at 90°C for 1 minute. Take care that no air
bubbles are trapped against the components.
2. Air dry out of direct sunlight.
High level thermal disinfection equivalent water bath parameters:
Process Temperature Process Time
1.
70 °C
100 minutes
2.
75 °C
30 minutes
3.
80 °C
10 minutes
4.
90 °C
1 minute
ResMed water tubs have been tested for disinfection for 100 cycles using hot water approximately 93°C for 10
minutes.

Other manuals for ResMed H5i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed H5i and is the answer not in the manual?

ResMed H5i Specifications

General IconGeneral
Water Capacity380 mL
Heated HumidificationYes
CompatibilityResMed S9 series CPAP machines
Power SourceIntegrated with CPAP device
Power Requirements24V via S9 device

Related product manuals