10
Viewing the alarms
To view the alarm list:
1. From the Monitoring screen, turn the dial clockwise until the last
Monitoring screen is displayed.
2. To view the alarm details, highlight the alarm and press the dial.
Testing the alarms
The LED alarm indicator will flash and the alarm will sound when power is connected to the device.
The alarms should be tested weekly to ensure they are working correctly. Follow the procedures in
this section to test the alarms. When completed, press Start/Stop and return the device to the
settings appropriate to the patient.
Setup for testing the alarms:
Turn off all adjustable alarms.
Set up the device with the air tubing attached, but no mask.
Set Ramp and SmartStart to Off .
To test the Tube disconnect alarm:
1. Disconnect air tubing at the air outlet.
2. Press Start/Stop.
The alarm activates after 510 seconds.
To test the High leak alarm:
1. Set the High leak alarm to On.
2. Leave the open end of the air tubing unblocked.
3. Press Start/Stop.
The alarm activates after 1030 seconds.
Caring for your device
It is important that you regularly clean your Lumis device to make sure you receive optimal therapy.
The following sections will help you with disassembling, cleaning, checking and reassembling your
device.
WARNING
Regularly clean your tubing assembly, humidifier and mask to receive optimal therapy and to
prevent the growth of germs that can adversely affect your health.

Other manuals for ResMed lumis series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed lumis series and is the answer not in the manual?

ResMed lumis series Specifications

General IconGeneral
Device TypeNon-invasive ventilator
Pressure Range4-25 cm H2O
ModesCPAP, S, ST, T, iVAPS
Ramp Time0-45 minutes
HumidificationIntegrated heated humidifier
Data StorageSD card
Power SupplyAC 100-240V, 50-60Hz
Weight1.3 kg
Operating AltitudeUp to 2, 591 m
Heated TubingCompatible with ClimateLineAir heated tubing
ConnectivityBluetooth
DisplayLCD

Related product manuals