Humidifier performance
The following settings have been tested at 22°C, 40% RH ambient conditions:
Mask pressure
cm H
2
O
RH output % Nominal system output AH
a
, BTPS
b
Setting 3 Setting 6 Setting 3 Setting 6
3 90 100 10 18
10 95 100 11.5 21
20 95 100 11 18
25 100 100 12 13.5
a. AH - Absolute Humidity in mg/L.
b. BTPS - Body Temperature Pressure Saturated.
Symbols
The following symbols may appear on your product or packaging.
Caution; Read instructions before use; Follow instructions for use; Drip proof;
Not drip proof; Type BF equipment; Class II equipment; Start/Stop;
Manufacturer; European RoHS; Lot number; Part number; Serial number;
Direct current; Not for use on more than one patient.; Maximum water level;
Disinfectable up to 200°F (93°C); Use distilled or deionized water only; Lock/unlock;
Remove tub to fill; R
x
Only Prescription only (In the US, Federal law restricts these
devices to sale by or on the order of a physician.); China pollution control logo 1; China
pollution control logo 2.
Servicing
The S9 and H5i devices are intended to provide safe and reliable operation when operated in
accordance with the instructions provided by ResMed. ResMed recommends that the S9 and H5i be
inspected and serviced by an authorized ResMed Service Center if there is any sign of wear or
concern with device function. Otherwise, service and inspection of the devices generally should not
be required during the five year design life of the device.