Pathology defaults
Setting Mode Obstructive
lung
disease
Restrictive
lung
disease
Obesity
Hypoventilation
Syndrome
Normal
lung
mechanics
S
ST
T
PAC
iVAPS
IPAP [cm H
2
O] 13 11 15 11
EPAP or PEEP [cm H
2
O] 5 5 7 5
PS [cm H
2
O] 8 6 8 6
Rise time [ms]* 150 300 300 200
Fall time [ms]* 200 200 200 200
Trigger sensitivity Medium Medium Medium Medium
Cycle sensitivity High Low Medium Medium
Min PS [cm H
2
O] 4 4 4 2
Max PS [cm H
2
O] 20 20 18 20
Ti Min [sec] 0.3 0.5 0.5 0.5
Ti Max [sec] 1.0 1.5 1.5 1.5
*The Rise/Fall Time milliseconds scale is an approximate only.
Note: iVAPS mode is only available in Stellar 150.
See the Stellar Clinical Guide for full details.
Not all devices are available in all regions.
Quick Setup Guide
English
Adult and paediatric ventilators