8 January 2008 Retsch GmbH 10 Doc. no. D 98.025.9999
Maximum relative humidity 80% at temperatures up to 31°C, with a
decrease to 50% relative humidity at 40°C.
At high air humidity the electrical and mechanical components
can become damaged and performance data can change to an
unknown extent.
Installation height :
max. 2000 m above sea level
Installation
Install the AS 200 tap only on a stable, safe surface in order
to prevent the transmission of undesirable vibrations.
Alignment
To ensure that the AS 200 tap can stand securely, all 4 feet
must be in contact with the installation surface.
Electrical connection
The voltage and frequency for the AS 200 tap are
specified on the type plate.
Make sure that the levels correspond to the available
power supply system.
Connect the AS 200 tap to the power supply system with
the aid of the connection cable included in the scope of
supply.
When connecting the mains cable to the power supply, use
an external fuse in conformance to the regulations
applicable to the place of installation.
Non-adherence to the levels specified on the type plate can
damage electrical and mechanical components.