28
29
Net Charge @ 500Ω [0.001]mC
Maximum Phase Charge @ 500 Ω Foot: 18µC Body: 18µC
Maximum Current Density Foot: 0.011 mA/cm² Body: 0.084 mA/cm²
Maximum Power Density @ 500Ω
Foot: 0.68 mW/cm² Body: 1.66 mW/cm²
Burst Mode: a. Pulses per burst Foot: 32-328 Body: 32-328
b. Bursts per second
Foot: 0.11 - 0.24 Body: 0.1 - 0.2
c. Burst duration
Foot: 3.2 - 8.3s Body: 3.2 - 8.3s
d. Duty Cycle Foot: 0.75 – 0.89 Body: 0.75 – 0.89
ON Time Foot: 1.9 - 8.3s Body: 1.9 - 6.3s
OFF Time
Foot: 1.0 - 1.50s Body: 1.0s
Disposal: At the end of product life do not throw this product into normal household garbage, but take it to a
collection point for the recycling of electronic equipment.
Technical Specications & Symbols
Contraindications, Warnings and Cautions
Make sure you understand these before using the Revitive
Power
Time Remaining
Intensity Level
Center Positive Polarity
Class II medical electrical equipment double insulated
FCC mark
Certification mark employed on electronic products manufactured or sold in
the United States which certifies that the electromagnetic interference from
Revitive is under limits approved by the Federal Communications Commission
Technical Specications
Name of product Revitive Essential
Model RLV
Weight 1.13kg
Dimensions
Ø 355 mm x 77(D) mm
Power Consumption 5W
Power Source:
AC adapter – UL (Underwriters Laboratories) Safety
Mark
Input
Output
100-240V AC, 50/60Hz, 0.18A
5v DC, 1.0A
Output Specifications:
Waveform Monophasic
Shape Square symmetrical with polarity reversal
Maximum Output Voltage (+/- 15%) : @500Ω Foot: 21.5Vp Body: 21.5Vp
@2kΩ
Foot: 55Vp Body: 62.5Vp
@10kΩ
Foot: 65Vp Body: 85Vp
Maximum Output Current: (+/- 15%) : @ 500Ω Foot: 43mA Body: 43mA
@2kΩ
Foot: 27.5mA Body: 31.3mA
@10kΩ
Foot: 6.5mA Body: 8.5mA
Pulse Duration (+/- 10%) :
Foot: 450 or 970µs (Mode 7) Body 455µs
Frequency (+/- 10%) :
Foot: 20 to 44Hz, Body: 20 to 44Hz
2452_IFU06_18063164.indd 28-29
2452_IFU06_18063164.indd 28-29
19/05/2021 10:22