Description of the P2 Oxygen Concentrator
16
Alerts Table
Alarm item
Alarm condition
System process
Screen Display
Battery
Exhausted
Battery cycles > 500
or battery health <
50%
Alarm only
Battery is
exhausted.
Replace battery
Replace Sieve
Bed
Sieve bed expired or
Sieve bed chip error
Alarm only
Replace Sieve
Bed
Please contact
provider
Low Input
Voltage
Adapter input
< 17.0v
Auto-switch to
battery until the
adapter input >
18v
Low Input Voltage
Please check
adapter
Absence of
Breath
No breath detected
continuously for
more than 15
seconds
Alarm only
Absence of Breath
Please check
cannula
Oxygen
concentration
<87%
Concentration < 87%
continuously for
more than 300
seconds
Alarm only
Low O
2
:< 87%
Please contact
provider
Low Battery
5%RSOC20%
without adapter
Alarm only
Low Battery
Please charge
Oxygen
concentration
<50%
Concentration < 50%
continuously for
more than 300
seconds
Auto-shut down
after 30 seconds
Low O
2
:< 50%
Please contact
provider
Breath Sensor
Fail
Breath Sensor failed
Auto-shut down
after 30 seconds
Breath Sensor
failed
Please contact
provider