Operating Instructions
Ⅲ. OPERATING INSTRUCTIONS
General Operation
1. Keep the P2 in a well-ventilated area
Be sure to keep the intake and exhaust areas of the device clear of obstructions
to allow for proper air flow. Find a location for the P2 such that auditory alarms
can be heard. Turn off device when not in use.
Do not use P2 in the presence of flammable
anesthetics, detergents or other chemical vapors.
Do not block the air intake or air exhaust when operating
the equipment.
Blocked air circulation or close proximity to a heat source
can cause internal heat build-up. This can cause the device
to shut down, or damage to the concentrator.
The P2 Concentrator is designed for continuous use. It is
useful to operate the product frequently for optimal sieve
bed life.
P2 is shipped from factory with battery removed.
2. Ensure
the particle filter is in place.
Do not operate the
P2
without the intake filter installed.
Inhalation of system particles can damage the device.