EasyManua.ls Logo

RHYTHM P2 - Device Classifications

RHYTHM P2
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Specifications
31
Concentrator Specification--Continued
Power:
AC Power supply
Rechargeable Battery
AC Input: 100 to 240VAC 50 to 60 Hz
Voltage: 14.4VDC Rated capacity:6.8Ah
Battery Duration
Up to 5 hours
Battery Charging Time
Not more than 4 hours
Environmental
Ranges
Intended for
Operation
Temperature: 41 to 104˚F (5 to 40˚C)
Humidity: 10% to 90%, non-condensing
Altitude: 0 to 10,000 ft. (0 to 3048 meters,70kPa to 106
kPa)
Environmental
Ranges
Intended for Shipping
and
Storage
Temperature: -4 to 158˚F (-20 to 70˚C)
Humidity: 5% to 90%, non-condensing
Store in a dry environment
Altitude: 0 to 10,000 ft (0 to 3048 meters,70kPa to 106
kPa)
Transportation
Keep dry, Handle with care
Classifications
Mode of Operation:
Continuous Duty
Type of Protection
Against Electrical
Shock:
Class II
Degree of Protection
to Concentrator
Components Against
Electrical Shock:
Type BF
Not intended for cardiac application
Degree of Protection
to Concentrator
Components Against
Ingress of Water
IP22 - Not protected from dripping water.
Protected against ingress of solid objects > 12.5
mm.