Operating Instructions
7. Set to the flow rate prescribed by your physician or clinician
Press the + or buttons to adjust the
P2 to the desired flow rate. The current setting
can be viewed on the display and ranges from 1 to 5.
CAUTION
Make sure the power supply is located in a well-ventilated
area. During operation, the power supply may get hot. Make
sure the power supply is cool before handling.
8.
Wear the nasal cannula on your face and breathe through your nose.
The P2 will sense if you are breathing from it. If you are not
yet breathing through the cannula, the P2 will begin to pulse
automatically about once every 3 seconds. As soon as you
begin breathing through the cannula, the device will begin
delivering pulses based on your breathing. As your
breathing rate changes, the P2 will sense these changes
and adjust the amount of oxygen at the next inhale.
WARNING
If you feel any discomfort when using the device, consult
your doctor immediately
CAUTION
A "Low O2< 87%" alert will display on the screen if the
oxygen level drops below recommended levels. If the alarm
persists, contact your device provider.
CAUTION
The display will dim if there is no operation for 30 seconds.
You can press any button to light up the display.
CAUTION
The P2 will notify you with audible alert (if enabled) and a
display showing "Absence Of Breath" if no breath has
been detected for 15 seconds. After 15 seconds, the device
will enter auto-pulse mode until breath is detected. Once
breath is detected, the device will resume normal delivery
of oxygen.
General Information
To disconnect power, unplug the power cord from the wall outlet and disconnect it
from the P2.