System Specifications
34
EMC Information---Continued
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
6 Vrms Amateur
Radio & ISM
Bands between
150 kHz and 80
MHz
10 V/m
80 MHz to 2.7
GHz
3 Vrms
150 kHz to 80
MHz
6 Vrms Amateur
Radio & ISM
Bands between
150 kHz and 80
MHz
10 V/m
Portable and mobile RF
communications equipment
should not be used near any
part of the device, including
cables. Recommended
distance is 30 cm (or 1').
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Guidance and Manufacturers Declaration –Electromagnetic Emissions:
The concentrator is intended for use in the electromagnetic environment specified
below. The user of the concentrator should make sure it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environment -
Guidance
RF Emissions
CISPR 11
Group 1
The device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage
Fluctuations/Flicker
Emissions IEC
61000-3-3
Complies