RIGOL
Quick Guide for DG4000
12
CAUTION
Overvoltage protection of the output channel will take effect once any of
the following conditions is met.
Amplitude setting in the generator is greater than 4 Vpp; the input
voltage is greater than ±11.25 V (±0.1 V) and frequency is lower
than 10kHz.
Amplitude setting in the generator is lower than or equal to 4 Vpp;
the input voltage is greater than ±4.5 V 0.1 V) and frequency is
lower than 10kHz.
The message “OverLoad protect, The output is off! will appear on
the screen when overvoltage protection takes effect.