Chapter 2 Performance Verification Test RIGOL
DG4000 Performance Verification Guide
2-1
Chapter 2 Performance Verification Test
Taking DG4162 as an example, this chapter introduces the method of performance
verification test for DG4000 series function/arbitrary waveform generator.
Test items include:
Frequency Accuracy Test
AC Amplitude Accuracy Test
DC Offset Accuracy Test
AC Flatness Test
Harmonic Distortion Test
Rise/Fall Time Test
Spurious Signal Test
Overshoot Test
Note:
1) Please make sure the instrument has been warmed up for at least 30 minutes
under the specified operational temperature (18 to 28) before the test.
2) Any of the following tests must be done under the specified operational
temperature (18 to 28).
3) Please reset the instrument to the factory setting before or after executing any
of the following tests.