EasyManua.ls Logo

Rigol DG4202

Rigol DG4202
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Appendix
2 DG4000 Performance Verification Guide
AC Amplitude Accuracy Test
Channel: CH1
Setting
Amplitude
Settings
Measurement
Values
Limits calculated from
specification
[1]
Pass/Fail
20 mVpp
Frequency:
1 kHz
Offset: 0 V
DC
Impedance:
50 Ω
6.3 mVrms~7.9 mVrms
100 mVpp 34.3 mVrms~36.4 mVrms
500 mVpp 174.3 mVrms~179.3 mVrms
1 Vpp 349.4 mVrms~357.9 mVrms
5 Vpp 1.7496 Vrms~1.7864 Vrms
10 Vpp 3.5 Vrms~3.5721 Vrms
Channel: CH2
Setting
Amplitud
e
Settings
Measuremen
t Values
Limits calculated from
specification
[1]
Pass/Fail
20 mVpp
Frequency:
1 kHz
Offset: 0 V
DC
Impedance:
50 Ω
6.3 mVrms~7.9 mVrms
100 mVpp 34.3 mVrms~36.4 mVrms
500 mVpp 174.3 mVrms~179.3 mVrms
1 Vpp 349.4 mVrms~357.9 mVrms
5 Vpp 1.7496 Vrms~1.7864 Vrms
10 Vpp 3.5 Vrms~3.5721 Vrms
Note
[1]
: Typical (1 kHz Sine, 0 V
DC
Offset, >10 mVpp, Auto) ±1% of setting ±2 mVpp

Other manuals for Rigol DG4202