Chapter 2 Front Panel Operations  RIGOL 
DG800 User's Guide  2-51 
To Select Modulation Source 
 
Tap the Source parameter selection field to select an "Internal" or "External" 
modulation source. 
 
1.  Internal Source 
When the internal source is selected, the modulating waveform is set as a Square 
with 50% duty cycle. At this time, the rate at which the output phase "shifts" 
between "carrier phase" and "modulating phase" is determined by the 
modulation rate. 
 
2.  External Source 
When the external source is selected, the generator receives the external 
modulating signal from the rear-panel [CH1/Sync/Ext Mod/Trig/FSK] 
connector. 
 
Note: The [CH1/Sync/Ext Mod/Trig/FSK] connector controlling PSK 
modulation externally is different from controlling AM/FM/PM modulations 
externally. While controlling the PSK modulation, you can set the polarity. 
 
 
To Set Modulating Rate 
 
When the internal source is selected, tap the Mod.Rate parameter input field and 
use the numeric keypad to set the rate at which the output phase shifts between 
"carrier phase" and "modulating phase". 
  You can also use the arrow keys and the knob to input a desired frequency 
value. 
  The frequency range is from 2 mHz to 1 MHz, and the default is 100 Hz. 
Note: When the external modulation source is selected, this menu will be grayed out 
and disabled. 
 
 
To Set Modulating Phase 
 
Modulating phase indicates the phase of the modulating waveform. Tap the Phase 
parameter input field and use the numeric keypad to set the modulation phase. 
  You can also use the arrow keys and the knob to input a desired phase value. 
  The phase range is from 0° to 360°, and the default is 180°.