EasyManuals Logo

Rigol DL3000 Manual

Rigol DL3000
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
RIGOL Chapter 2 Specifications Test
DL3000 Performance Verification Manual
2-6
b) Turn on the digital multimeter. Select the DC voltage (DCV)
measurement function and set the range to "Auto". Make sure that the
electronic load works in CC mode. Read the digital multimeter and
record the reading (U
2
) of the digital multimeter. Calculate the current
value (I
3
= U
2
/R
M
).
c) Send the :MEAS:CURR? command via the remote interface, read and
record the returned current value (I
4
) of the query command.
d) Calculate the CC programming accuracy (|I
3
- I
0
|) and the CC readback
accuracy (|I
4
- I
3
|). Compare them with the specified specifications in
Table 2-4. Wherein, I
0
is the rated current value of the input channel.
7. Send the following command via the remote interface to disable the channel
input.
:SOUR:INP:STAT 0
/*Disables the channel input.*/

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals