EasyManua.ls Logo

Rigol DP831A - Page 43

Rigol DP831A
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 Constant Voltage Tests RIGOL
DP800 Performance Verification Manual
2-25
Table 2-24 DP832 CV Programming and Readback Accuracy Test Record Form
Channel CH1 (30V/3A) CH2 (30V/3A) CH3 (5V/3A)
Specification
Programming <25mV <25mV <12.5mV
Readback <20mV <20mV <7.5mV
The output voltage is 0
DP832 Setting
Voltage 0V 0V 0V
Current 3.2A 3.2A 3.2A
Measurement
Result
U
1
U
2
Programming:
|U
1
-0|
Readback:
|U
2
-U
1
|
The output voltage is U
R
[1]
DP832 Setting
Voltage 30V 30V 5V
Current 3.2A 3.2A 3.2A
Measurement
Result
U
3
U
4
Programming:
|U
3
-U
R
[1]
|
Readback:
|U
4
-U
3
|
Note
[1]
: The rated voltage of the specified channel.