RIGOL Chapter 2 Constant Voltage Tests
DP800 Performance Verification Manual
2-26
Table 2-25 DP821A CV Programming and Readback Accuracy Test Record Form
Channel CH1 (60V/1A) CH2 (8V/10A)
Specification
Programming <85mV <14mV
Readback <85mV <9mV
The output voltage is 0
DP821A
Setting
Voltage 0V 0V
Current 1.05A 10.5A
Measurement
Result
U
1
U
2
Programming:
|U
1
-0|
Readback:
|U
2
-U
1
|
The output voltage is U
R
[1]
DP821A
Setting
Voltage 60V 8V
Current 1.05A 10.5A
Measurement
Result
U
1
U
2
Programming:
|U
3
-U
R
[1]
|
Readback:
|U
4
-U
3
|
Note
[1]
: The rated voltage of the specified channel.