Chapter 2 Constant Voltage Tests RIGOL
DP800 Performance Verification Manual
2-27
Table 2-26 DP811A CV Programming and Readback Accuracy Test Record Form
Range Range1 (20V/10A) Range2 (40V/5A)
Specification
Programming <20mV <30mV
Readback <20mV <30mV
The output voltage is 0
DP811A
Setting
Voltage 0V 0V
Current 10.5A 5.3A
Measurement
Result
U
1
U
2
Programming:
|U
1
-0|
Readback:
|U
2
-U
1
|
The output voltage is U
R
[1]
DP811A
Setting
Voltage 20V 40V
Current 10.5A 5.3A
Measurement
Result
U
1
U
2
Programming:
|U
3
-U
R
[1]
|
Readback:
|U
4
-U
3
|
Note
[1]
: The rated voltage of the specified range.