RIGOL Chapter 2 Performance Verification Test
DS1000Z-E Performance Verification Guide
2-12
adjust the horizontal position and vertical position properly.
6) Rotate TRIGGER
LEVEL to adjust the trigger level to the middle of the screen.
5. Observe the screen of the oscilloscope. Press Cursor Mode "Manual" to turn on the manual
cursor function. Measure the zero point offset using manual cursor measurement and record the
measurement result.
6. Output a fast edge signal with 500 ps rise time and 3 V amplitude via Fluke 9500B.
7. Keep the other settings of the oscilloscope unchanged and set the vertical scale to 500 mV/div.
8. Repeat Step 5; measure the zero point offset and record the measurement result.
9. Turn off CH1. Test CH2 according to the method above and record the test results.
Test Record Form
Channel
Fast Edge Signal
Amplitude
Vertical
Scale
Test Result Limit
Pass/Fail
CH1
1.2 Vpp
200 mV/div
2.5 ns
3 Vpp
500 mV/div
CH2
1.2 Vpp
200 mV/div
3 Vpp
500 mV/div