RIGOL Appendix Test Record Form
DS1000Z-E Performance Verification Guide
2
Bandwidth Limit Test
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vrms1 Vrms2
Amplitude
Loss
[1]
CH1
100 mV/div
-3 dB to 3 dB
500 mV/div
CH2
100 mV/div
500 mV/div
Note
[1]
: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).
Time Base Accuracy Test
Channel Test Result ΔT
Calculation
Result
[1]
Limit Pass/Fail
CH1
±(25 ppm + 5 ppm/year × number of years
that the instrument has been used
[2]
)
Note
[1]
: Calculation Result = Test Result ΔT/1 ms.
Note
[2]
: For the number of years that the instrument has been used, please calculate according to the date in the
verification certificate provided when the instrument leaves factory.
Zero Point Offset Test
Channel
Fast Edge Signal
Amplitude
Vertical Scale Test Result Limit
Pass/Fail
CH1
1.2 Vpp
200 mV/div
2.5 ns
3 Vpp
500 mV/div
CH2
1.2 Vpp
200 mV/div
3 Vpp
500 mV/div