Programming Introduction                                                                                                                RIGOL 
                                                                     
Programming Guide for DS1000CA Series 
Function Explanation: 
The commands are used to set and query the vertical offset of the waveform. 
<n> is 1 or 2. 
When Scale>200mV, the range of <offset> is -40V~+40V; 
When Scale≤200mV, the range of <offset> is -800mV~+800mV. 
 
Returned Format: 
Query returns the value of <offset> in V. 
 
Example: 
:CHAN2:OFFS 20    Set the vertical offset of channel 2 as 20V 
:CHAN2:OFFS?      Query returns 2.000e+01 
 
 
6.  :CHANnel<n>:PROBe 
Command Format: 
:CHANnel<n>:PROBe <attn> 
:CHANnel<n>:PROBe? 
 
Function Explanation: 
The commands are used to set and query the attenuation factor of the probe 
to keep the gauge exact. The options of <attn> are 1, 5, 10, 50, 100, 500 or 
1000. <n> is 1 or 2. 
 
Returned Format: 
Query returns the set value of <attn> in scientific numeric notation. 
 
Example: 
:CHAN2:PROB 10    Set probe attenuation of channel 2 as 10X   
:CHAN2:PROB?     Query returns 1.000e+01 
 
 
7.  :CHANnel<n>:SCALe 
Command Format:   
:CHANnel<n>:SCALe <range> 
:CHANnel<n>:SCALe?