No. of parameters 2
Save changes With TDD1
Increment Trade Counter Yes
Parameter Details
Parameter Description Range Default
1 Weighing mode 1 = Single range
2 = Dual range
3 = Dual interval
1
2 Trade mode 0 = OIML
1 = INDUST
2 = NTEP
1
Use the WMD command to setup the weighing mode of the instrument. This setting is a fundamental scale
build parameter and should be used along with the IAD and ICR commands before the unit is calibrated.
Example
Command Answer Description
S01; Select unit 1
WMD?; 1,0 CRLF Query current weighing mode
WMD2,1; 0 CRLF Change to dual range, industrial mode
WMD?; 2,1 CRLF Weighing mode is dual range, industrial
TDD1; 0 CRLF Save settings
21.3.57 ZST: Zero Settings
Set the various options associated with zero balance.
General
No. of parameters 4
Save changes With TDD1
Increment Trade Counter Depends on parameter
Parameter Details
Parameter Description Range Default Trade Counter
1 Zero on startup 0 = Off
1 = On
0 No
2 Zero tracking 0 = Off
1 = 0.5d in 1sec
2 = 1.0d in 1 sec
3 = 2.0d in 1 sec
4 = 5.0d in 1 sec
5 = 0.5d in 0.5 sec
6 = 1.0d in 0.5 sec
7 = 2.0d in 0.5 sec
8 = 5.0d in 0.5 sec
9 = 0.5d in 0.2 sec
10 = 1.0d in 0.2 sec
11 = 2.0d in 0.2 sec
12 = 5.0d in 0.2 sec
0 Yes
3 Zero setting range 1 = -20% .. 20%
2 = -100% .. 100%
3 = -2% .. 2%
4 = -1% .. 3%
3 Yes
4 Zero dead band 0 .. 100000 0 Yes
159 C500-600-2.0.3

Table of Contents

Other manuals for Rinstrum C510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rinstrum C510 and is the answer not in the manual?

Rinstrum C510 Specifications

General IconGeneral
BrandRinstrum
ModelC510
CategoryTouch Panel
LanguageEnglish

Related product manuals