Specifications • 7
Specifications
Table 1 lists specifications for the Beacon 200 Gas Monitor.
CAUTION: Cleaning is not to be performed.
CAUTION: The operator of this instrument is advised that if the equipment
is used in a manner not specified in this manual, the protection
provided by the equipment may be impaired.
Table 1: Beacon 200 Specifications
Description Specification
Input Power
115/220V ~ ±10%, 50/60Hz, 0.5/0.3A
or
24 V +10% -8%, 0.6A VDC
Construction (housing) Fiberglass/polyester with lexan window (NEMA 4X)
Dimensions 10.5 in. H x 8.5 in. W x 6.25 in. D
(267 mm H x 216 mm W x 158 mm D)
Weight 8 lbs.
Environmental Conditions • For indoor or outdoor locations (Type 4X)
• 2000m max. altitude
• -20°C to 50°C (-4°F to 122°F) max. ambient
• Maximum humidity of 80% relative
• Mains supply voltage fluctuations not exceeding ± 10% of nominal
• DC supply voltage fluctuations not exceeding +10% -8% of nominal
• Overvoltage Category II, Pollution Degree 2
Safety/Regulatory
C US
186718
Relays • Relay contacts rated for 10A @ 115/220V~ resistive or 10A @ 30V
resistive
• SPDT Form C
User Controls • Reset switch
• Program buttons: ESCAPE, UP/YES, DOWN/NO, and ENTER
Standard Accessory Operator’s manual (this document)