.1
59 / 93
7.3.2 Database Factory Reset
Request the database initialization to factory default from the other device. It is sent by the
host.
The system can only be reset in ‘Standby’ mode.
Once such a command is received it will be confirmed an EAR and will then change to the
‘error’ state immediately – a restart is required to make the reset active - the database is re-
placed with the factory provided during startup only.
7.3.2.1 Automated equipment command (EAC - event U07)
General: This message is used to send equipment commands from one application to
another (e.g., automated Equipment to a Laboratory Automation System).
cobas c 111 instrument: The message will trigger the initialization of the database at the
next system startup. See segment description for specifics.
EAC^U07 Equipment Command Message
H Message Header
M|1|EQU Equipment Detail
M|1|ECD Equipment Command Detail – see segment information
L
7.3.2.2 Automated equipment response (EAR - event U08)
This message is used to send equipment responses to previously issued commands from
one application to another (e.g., automated Equipment to a Laboratory Automation System).
EAR^U08 Equipment Command Message
H Message Header
M|1|EQU Equipment Detail
M|1|ECD Equipment Command Detail – reply command reference number as received
M|1|ECR Equipment Command Response OK or UN if the command cannot be accepted
See segment information for details
L