EasyManuals Logo

Roche Cobas C111 User Manual

Roche Cobas C111
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
cobas c 111
Host Interface Manual
Version
2
.1
39 / 93
The Manufacturer Specific Records are specific for the cobas c 111 instrument and must
be received by the host
Level Record Name Identifier
1 Photometry Calibration Result M.CR
1 ISE Calibration Result M.ICR
3 Photometric Raw Data M.RR
1 Reagent Disk Loading INR^U06
INU^U05
1 Database Factory Reset (Command) EAC^U07
EAR^U08
Table 9: Manufacturer Defined Record Types and Levels
7.2.2 Record Description
7.2.2.1 Field Attributes
Types of attributes held by a field are explained below.
No. Attribute Description
1 Field Name Name of the relevant field.
Fields printed with underlined-italic font are mandatory fields.
2 Reference Position of the field. Order in which the relevant fields appear in a record.
3 Format The format for a field is one of the following:
ST: String: A character string
TX: Text: A group of character strings that can be printed at the terminal. It is an
optional character string; however, a special escape sequence is defined for a display at the
terminal.
NM: Numeric: A numeric value. Positive (+) or negative (-) is indicated before the numeric
value. If it is not indicated, it is treated as positive (+).
If a decimal point is not included, the numeric value is treated as an integer. There are no
restrictions for placing ‘0’ in the front and for ‘0s’ placed at the end of numbers with decimal
points.
For certain fields where the resolution with a decimal number cannot be achieved a scientific
format is used i.e. (1.234E03)
DT: Date. Always use the 4-digit Christian year. The format is YYYYMMDD (YYYY is
the 4-digit Christian year, MM is the month, DD is the day). For example, September 5, 1998 is
indicated 19980905.
TM: Time: 24h format (corresponds with the U.S. Military time)
The format is HHMMSS (HH is the hour, MM is the minute, and SS is the second).
TS: Time Stamp. A combination of DT and TM.
The format is: YYYYMMDDHHMMSS.
CM: Combination. A field in which multiple data are combined by a component
delimiter.
4 Comments
(Data Content)
Contains field contents and any relevant field comments.
Note: A field with the message ‘Field does not contain data‘ in the column means that even
though it is provided by ASTM, it may be ignored when received.
Table 10: Field Attributes