EAR^U08 Reports the reception of the command or that
the request has been ignored
Table 5: Messages Transmitted by the Analyzer
6.4.1 Messages received: Host Analyzer
Messages transmitted by the host are shown in the table below (messages received by the
analyzer).
Messages Syntax / Records Comment or Special
Instructions
Reasons
Analytical Data
Transmission
Request Sample
H
Q|n|^SID
L
RSREQ ^ REAL Makes inquiry for the results of a test.
SID=SampleID, Test ID is assumed as ALL
System replies with RSUPL.
Analytical Data
Transmission
Request Cal
H
Q|n|||AppC
ode
L
RSREQ ^ REAL Makes inquiry for the results of a calibration
Only one appCode allowed
The system replies with PCUPL or ICUPL .
Note: the system reports calibration results of
the currently active calibration if still available –
receive the real time uploads to avoid gaps in
the reporting sequence.
Test Orders H
P|n
O|1|SID…
[C|1|..]
L
TSDWN ^ REPLY
TSDWN ^ BATCH
Orders test request for a sample as answer to a
Test Selection Inquiry
Orders test request for a sample by instruction
at the host
Reagent Tray
Loading Inquiry
H
M.EQU
M.INV
L
INR^U06 Makes an inquiry for the reagent tray loading
cobas c 111 instrument replies with INU
Database factory
reset request
H
M.EQU
M.ECD
L
EAC^U07 Factory command to reset the database to its
default status (reloads the empty database
stored on the system)
cobas c 111 instrument replies with EAR
This is for manufacturer’s exclusive use and is
only included here for completeness. There is no
use of this feature for customers.
Table 6: Messages Transmitted by the Host