7.4.4 Equipment command response segment (ECR)
The equipment command response segment contains the receiving component’s response
to the previously received command.
# Field Name
Data
Type
Opt
In
Opt
Out
Re-
peat
Description
1 Record Type ID Char R R M : Manufacturer Record
Instrument transmits upper case characters and
receives upper or lower case.
2 Sequence Number Pos_int R R 1 : Sequence number is always one (1).
3 Record type sub ID Text
^text
^text
^text
R
R
O
O
O
cobas c 111 instrument receives data: field is
ignored
cobas c 111 instrument transmits data:
Format: |ECR ^RO^ c111 ^1|
ECR
^RO (Roche specific record identifier)
^Instrument abbreviation (c111)
^Version# of protocol implementation (1)
4 Command Response CE R Format: |OK|
cobas c 111 instrument transmits data: only the first
field
<identifier (ST)> is used according to the table below
only “OK” and “UN” is supported see parameters
5 Date/Time Completed TS
Not used / ignored by cobas c 111 instrument
6 Command Response
Parameters
ST O Y Format: |C111 -2509|
cobas c 111 instrument transmits data: only one field
(ST9)
Reports the same tag to confirm the reset or an empty
field to indicate that the reset request was ignored
‘C111 -2509’ a string field which has to match exactly.
EndOfRecord
7.4.4.1 Command response (CE)
Components: <identifier (ST)> ^ <text (ST)> ^ <name of coding system (IS)> ^ <alternate
identifier (ST)> ^ <alternate text (ST)> ^ <name of alternate coding system
(ST)>
Definition: This field identifies the response of the previously issued command. Refer to
HL7 table 0387 – Command response for valid values.
Table 0387 - Command response
Value Description
OK * Command completed successfully
UN * Command cannot be completed for unknown reasons
7.4.4.2 Command response parameters (ST)
Definition: This field identifies any associated parameters that relate to the returned
response command message.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cobas C111 and is the answer not in the manual?

Roche Cobas C111 Specifications

General IconGeneral
BrandRoche
ModelCobas C111
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals