cobas c 111
Host Interface Manual
Version
2
.1
51 / 93
7.2.2.8 Comment Record (following the result record)
Record to transfer comments regarding the analytical data. This record is transmitted by the
analyzer only; it is not transmitted by the host (i.e. ignored when a cobas c 111 instrument
receives such a record).
C|1|I|43|I{RT}
Field Name Reference Format
Comments (Data Content) C|11.1.
Record Type ID 11.1.01 ST ‘C’
Sequence Number 11.1.02 NM Indicates the sequence number of the Comment Record in the current
layer. Since this record is in the layer following the Test Request Record, it
is reset to ‘1’ each time a new Test Request Record is presented and then
numbered consecutively; 1, 2, etc.
Comment Source 11.1.03 ST Indicates the source of the Comment.
Currently, this record is issued only from the analyzer, so ‘I’ is always used.
Comment Text:
Flag Code
^Flag Comment
11.1.04 NM
repeated
A comment record is added following a flagged result.
All activated flags are transmitted.
Repeat delimiters are used if more than one flag is activated.
Comment Type 11.1.05 ST Character limiting the comment format.
This record is issued only from the analyzer, so ‘I’ is always used.
EndOfRecord

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cobas C111 and is the answer not in the manual?

Roche Cobas C111 Specifications

General IconGeneral
BrandRoche
ModelCobas C111
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals