Maintenance and Trouble Shooting
79
C
C
Grease the new O-rings as described in Section C.1.2.1 above.
Note: Lubricating the O-rings before placing them on the nozzles is important to ensure
correct reaction tip uptake and alignment.
Install new O-rings as shown below:
While holding the O-ring on the Exchange
Tool with your thumb, pull the entire O-
ring changer downward, away from the
nozzle.
The O-ring is now removed from the
nozzle:
Place a new, greased O-ring on the shaft
of the O-Ring Exchange Tool. Make sure
the O-ring is well below the pin.
Insert the pin of the O-Ring Exchange
Tool as deep as possible into the nozzle
hole:
Use your thumb and forefinger to push
the O-ring up the shaft onto the nozzle.
Continue pushing until the O-ring is
seated in the nozzle groove.
Slowly withdraw the Exchange Tool from
the nozzle.
7
8
1
2
3
User Maintenance
O-ring Maintenance
MagNA Pure Compact Manual.book Seite 79 Dienstag, 28. Februar 2006 4:16 16