16
Important Information regarding the MagNA Pure Compact Instrument Operator’s Manual
Additional Information
Chapter C:
1. User Maintenance
1.3 Create a Problem Report
For troubleshooting by a Roche representative, error messages and confi guration fi les can be
exported to a Problem Report.
The Problem Report is a zipped fi le containing log, confi guration and version fi les.
To create a Problem Report:
1
Insert an empty USB Memory Stick on the right side of the Instrument. On the “Main Menu”
screen, press the ‘Maintenance’ button. Press the ‘Create Problem Report’ button and select “OK”
in the pop-up window.
2
The Problem Report is created and stored on the attached USB Memory Stick. A pop-up window
will open when the Problem Report has been saved.
1. User Maintenance
1.3 Show All Protocols
The “Show All Protocols” option lists all the protocols which are installed on the MagNA
Pure Compact Instrument.
To “Show All Protocols”
1
On the “Main Menu” screen, press the ‘Maintenance’ button. Select the “Show All Protocols”
button.
2
All protocols that are installed on the Instrument will be displayed, as shown in the screenshot
below.
22051208 MagNA Pure Compact SW Update 1.1.2 Addendum 2.indd 1622051208 MagNA Pure Compact SW Update 1.1.2 Addendum 2.indd 16 24.03.2009 16:05:3024.03.2009 16:05:30