EasyManua.ls Logo

Rohde & Schwarz R&S FPL1000 - Page 22

Rohde & Schwarz R&S FPL1000
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing for Use
R&S
®
FPL1000
22Getting Started 1179.4657.02 ─ 10
To connect the AC power
1. Plug the AC power cable into the AC power connector on the rear panel of the
instrument. Only use the AC power cable delivered with the R&S FPL1000.
2. Plug the AC power cable into a power outlet with ground contact.
The required ratings are listed next to the AC power connector and in the data
sheet.
For details on the connector, refer to Chapter 5.2.1, "AC Power Supply Connec-
tion and Main Power Switch", on page 44.
4.5.2 Connecting an Optional DC Power Supply (R&S FPL1-
B30)
The R&S FPL1000 can also be equipped with an optional DC power supply con-
nector (R&S FPL1-B30). If installed, the R&S FPL1000 can be operated by a DC
voltage of +12 V to +24 V. For details on the connector see Chapter 5.2.2, "Li-Ion
Battery Packs and DC Power Connector", on page 45.
If you use an external power supply unit to supply safety extra-low DC voltage
(SELV) to the instrument, be sure to meet the requirements for reinforced/double
insulation in accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No.
1010.1) or DIN/EN/IEC 60950 (UL 1950, CSA C22.2 No. 950). Provide current
limitation in accordance with DIN EN 61010-1 Appendix F2.1. Use a cable no lon-
ger than 3 m.
Also see "Connecting to power" on page 9.
Connecting to Power

Related product manuals