R&S FSH Operation in Receiver Mode
1145.5973.12 2.21 E-11
Selecting the frequency step size:
The frequency resolution in the receiver mode is 100 Hz. The tuned step size can be changed as
required for the application
! Press the FREQ key.
! Press the FREQ STEPSIZE softkey.
! Set the required step size in the selection table.
! Confirm with the ENTER key.
! You can set any step size you want by using MANUAL....
! To do so, select MANUAL... for the step size in the selection table.
! Using the rotary knob or cursor keys, change the tuned step size and confirm with the ENTER key,
or, using the numeric keypad, manually enter a step size and confirm by pressing the units key.
Tuning the frequency in channel grids:
As an alternative to entering the frequency, the R&S FSH can also be tuned in channels. The channel
tables that the R&S FSH uses to set channel frequencies are defined either by using the R&S FSH
View software or by directly entering the first channel number, the associated frequency, the number of
channels and the channel spacing.
! Press the FREQ key.
! Press the CHANNEL MODE softkey.
The R&S FSH now uses the active channel table. The FREQ softkey for frequency entry is renamed to
CHANNEL for channel entry, and the R&S FSH displays the channel number rather than the frequency.
The channel numbers are now used to tune the frequency.
Selecting a channel table that was predefined using R&S FSH View:
! If the channel display is active (CHANNEL MODE
softkey active in the FREQ menu), press the
CHANNEL TABLE softkey.
The R&S FSH will display the stored channel tables.
! Using the rotary knob or cursor keys, select the
channel table you want.
! To activate the channel table, press the SELECT
softkey.