Operation in Receiver Mode R&S FSH
1145.5973.12 2.22 E-11
Direct entry of a channel table:
! If the channel display is active (CHANNEL MODE softkey active in the FREQ menu), press the
CHANNEL TABLE softkey.
! Press the SELECT USER TAB softkey.
The R&S FSH will use the last channel table that was entered directly.
! Press the DEFINE USER TAB softkey.
The R&S FSH will open the submenu for defining the
channel table.
! Press the DEFINE USER TAB softkey again.
! Enter the number of the first channel and confirm with the ENTER key.
! Press the DEFINE USER TAB softkey.
! Using the rotary knob or the cursor keys, select 1
ST
CHANNEL FREQ... from the menu and confirm
with the ENTER key.
! Enter the frequency for the first channel number.
! Press the DEFINE USER TAB softkey.
! Using the rotary knob or the cursor keys, select NO OF CHANNELS... from the menu and confirm
with the ENTER key.
! Enter the number of channels and confirm with the ENTER key.
! Press the DEFINE USER TAB softkey.
! Using the rotary knob or the cursor keys, select CHANNEL SPACING... from the menu and confirm
with the ENTER key.
! Enter the frequency spacing for the channels and confirm with the ENTER key.
! Press the EXIT key to exit the menu for defining channel tables.
The R&S FSH will now show channel numbers rather than the frequency. It also shows the associated
frequency above Channel.
Selecting the reference level:
The reference level is the maximum level of the analog bar-graph display. It must be set such that the
level display is located within the bar-graph scale.
! Press the REF LEVEL softkey in the main menu of the receiver mode (MEAS key). Or press the
AMPT key.
! Using the rotary knob or cursor keys, change the reference level or, using the numeric keys, enter a
new reference level.
! Confirm with the ENTER key.