6.8 R&S NRP2
With the R&S NRP power sensors and an R&S NRP2, you can measure power
with up to four power sensors simultaneously. All sensor-dependent measure-
ment functions can be used and the results can be displayed in parallel.
Required equipment
R&S NRP power sensor
R&S NRPZK6 cable to connect the sensor to the R&S NRP2
R&S NRP2 base unit with FW version 7.11 or higher
Setup
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 6-5: Setup with an R&S NRP2 base unit
1 = Signal source
2 = R&S NRP power sensor
3 = Host interface connector
4 = R&S NRPZK6 cable
5 = Sensor input connector of the R&S NRP2
6 = R&S NRP2 base unit
Incorrectly connecting/disconnecting the R&S NRP power sensors can
damage the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Chapter 4, "Preparing for Use", on page 10.
Starting a measurement
1. Connect the cables as shown in Figure 6-5:
a) Connect the R&S NRPZK6 cable to the host interface connector of the
sensor.
R&S NRP2