Preparing for Use
R&S
®
SMB100A
16Quick Start Guide 1407.0812.62 ─ 20
Left = Stacked correctly, same dimensions
Middle = Stacked correctly, different dimensions
Right = Stacked incorrectly, too many products
4.
NOTICE! Overheating can damage the product.
Prevent overheating as follows:
● Keep a minimum distance of 10 cm between the fan openings of the product
and any object in the vicinity.
● Do not place the product next to heat-generating equipment such as radiators
or other products.
4.4.2 Mounting the R&S SMB in a Rack
To prepare the rack
1. Observe the requirements and instructions in "Setting up the product" on page 8.
2.
NOTICE! Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S SMB in a rack
1. Use an adapter kit that fits the dimensions of the R&S SMB to prepare the instru-
ment for rack mounting.
a) Order the rack adapter kit designed for the R&S SMB. For the order number,
see the data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the R&S SMB to shelf height.
3. Push the R&S SMB onto the shelf until the rack brackets fit closely to the rack.
4. Tighten all screws at the rack brackets with a tightening torque of 1.2 Nm to secure
the R&S SMB in the rack.
Setting Up the R&S