R&S
®
SMBV100A
Vector Signal Generator
Operating Manual
Operating Manual
1176.8016.02 ─ 17
(;ÚÞ@2)

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Questions and Answers

  • S
    Shane PadillaAug 21, 2025
    What to do if Rohde & Schwarz Portable Generator shows 'Adjustment data invalid'?
    • A
      Amy BlanchardAug 21, 2025
      If the Rohde & Schwarz Portable Generator shows 'Adjustment data invalid', the adjustment data are invalid and must be restored. The adjustment data have to be generated again by an internal or external adjustment or loaded into the instrument.
  • N
    Nicole ChanAug 25, 2025
    What to do if Rohde & Schwarz Portable Generator says 'Adjustment data missing'?
    • S
      Scott SpencerAug 25, 2025
      If the Rohde & Schwarz Portable Generator indicates that adjustment data are missing, they have to be generated first by an internal or external adjustment or loaded into the instrument.
  • S
    Sheila MillerAug 26, 2025
    Why does Rohde & Schwarz Portable Generator display 'Extern reference out of range or disconnected'?
    • L
      logan35Aug 26, 2025
      The Rohde & Schwarz Portable Generator displays 'Extern reference out of range or disconnected' because the external reference is selected, but no external signal is applied, or the signal is out of range. To solve this: * Check the selected reference signal source (internal or external) in the 'Setup > Reference Oscillator' dialog. * Change the setting to 'internal' if no appropriate external source is available.
  • M
    Maria CummingsAug 28, 2025
    What to do if Rohde & Schwarz Portable Generator displays 'Adjustment failed'?
    • C
      chadpattersonAug 28, 2025
      If the Rohde & Schwarz Portable Generator displays 'Adjustment failed', it means the adjustment could not be executed. The adjustment data have to be generated first by an internal or external adjustment or to be loaded into the device.
  • A
    Ashley StaffordAug 30, 2025
    What to do if my Rohde & Schwarz Portable Generator displays 'File contains invalid data'?
    • L
      leemitchellAug 30, 2025
      If your Rohde & Schwarz Portable Generator displays 'File contains invalid data', check the file extension, as the selected file contains data that is not valid for the file type.
  • A
    Alex ReedSep 1, 2025
    What does it mean when Rohde & Schwarz Portable Generator displays 'Hardware revision out of date'?
    • H
      Holly ArcherSep 1, 2025
      When the Rohde & Schwarz Portable Generator displays 'Hardware revision out of date', it means a later version of certain parts of the instrument is necessary to execute the function selected because the driver does not support the installed version of a module.
  • A
    Andrew PayneSep 3, 2025
    What does it mean when Rohde & Schwarz Portable Generator displays 'Invalid filename extension'?
    • M
      michelle50Sep 3, 2025
      If the Rohde & Schwarz Portable Generator displays 'Invalid filename extension', check the file extension. For instance, the file extension for waveform list files is *.wv. It is not possible to enter another file extension when storing a list, as the file extension is not valid for the required operation.
  • S
    Sonia JohnsonSep 5, 2025
    What to do if Rohde & Schwarz Portable Generator displays 'Filename missing'?
    • D
      Dana StevensonSep 5, 2025
      If the Rohde & Schwarz Portable Generator displays 'Filename missing', a file name has to be entered when creating a new list because the required operation cannot be executed as the file name is not specified.
  • K
    Kimberly HuynhSep 7, 2025
    What does it mean if Rohde & Schwarz Portable Generator displays 'Unknown list type specified'?
    • A
      April EdwardsSep 7, 2025
      If the Rohde & Schwarz Portable Generator displays 'Unknown list type specified', check the selected list type because it is not valid for the required operation.
  • M
    marycampbellSep 9, 2025
    What to do if my Rohde & Schwarz Portable Generator displays 'No current list'?
    • C
      Crystal SilvaSep 10, 2025
      If the Rohde & Schwarz Portable Generator displays 'No current list', a list has to be selected in the related menu to execute the required operation. If no list is available, a new list must be created, because there is no list selected.