Remote Control Commands
R&S
®
SMBV100A
579Operating Manual 1176.8016.02 ─ 17
Furthermore the following suffixes denote:
●
Sensor assignment READ<ch> and INIate<hw>; range: [1] to 4
●
Sensor mapping list: ELEMent<ch>; range [1] to 25
:SENSe<ch>[:POWer]:APERture:DEFault:STATe............................................................. 579
:SENSe<ch>[:POWer]:APERture:TIMe.............................................................................579
:INITiate<ch>[:POWer]:CONTinuous................................................................................580
:READ<ch>[:POWer]?.................................................................................................... 580
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect.......................................................581
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe....................................................... 581
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?........................................................ 581
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe.......................................................... 582
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority.........................................................582
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?..................................................................582
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER]..................................................................583
:SENSe<ch>[:POWer]:FILTer:NSRatio.............................................................................583
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe..................................................................583
:SENSe<ch>[:POWer]:FILTer:SONCe..............................................................................584
:SENSe<ch>[:POWer]:FILTer:TYPE................................................................................ 584
:SENSe<ch>[:POWer]:LOGGing:STATe...........................................................................585
:SENSe<ch>[:POWer]:FREQuency..................................................................................585
:SENSe<ch>[:POWer]:OFFSet........................................................................................586
:SENSe<ch>[:POWer]:OFFSet:STATe............................................................................. 586
SENSe<ch>[:POWer]:SNUMber?.................................................................................... 586
:SENSe<ch>[:POWer]:SOURce.......................................................................................587
SENSe<ch>[:POWer]:STATus[:DEVice]?......................................................................... 587
SENSe<ch>[:POWer]:SVERsion?....................................................................................587
SENSe<ch>[:POWer]:TYPE?..........................................................................................588
:SENSe<ch>[:POWer]:ZERO.......................................................................................... 588
:SENSe<ch>:UNIT[:POWer]............................................................................................588
:SENSe<ch>[:POWer]:APERture:DEFault:STATe <UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command :SENSe<ch>[:POWer]:
APERture:TIMe on page 579.
Parameters:
<UseDefAp> 0 | 1 | OFF | ON
*RST: 1
Example:
SENS:POW:APER:DEF:STAT 0
deactivates the default aperture time of the sensor.
Manual operation: See "Use Default Aperture Time" on page 247
:SENSe<ch>[:POWer]:APERture:TIMe <ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
SENSe, READ and INITiate Subsystems