Remote Control Commands
R&S
®
SMBV100A
542Operating Manual 1176.8016.02 ─ 17
:BLER:SETup:TIMeout <Timeout>
:BERT:SETup:TIMeout <Timeout>
Sets the time out.
Parameters:
<Timeout> float
Range: 0.1 to 1
Increment: 0.001
*RST: 0.1
Manual operation: See "Timeout" on page 196
:BLER:STARt
:BERT:STARt
Starts a continuous measurement.
These commands automatically set BERT:SEQuence or BLER:SEQuence to AUTO
and BERT:STATe or BLER:STATe to ON.
Example:
BERT:STAR
starts a continuous measurement.
Usage: Event
Manual operation: See "Mode" on page 192
:BLER:STATe <State>
:BERT:STATe <State>
Switches the measurement on/off. Depending on the selected mode, either a continu-
ous measurement (BERT|BLER:SEQ AUTO) or a single measurement
(BERT|BLER:SEQ SING) is carried out. A single measurement must be triggered
(:TRIG:BERT|BLER).
These commands are automatically set to ON by BERT:STARt or BLER:STARt and to
OFF by BERT:STOP or BLER:STOP.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BERT:SEQ SING
selects a single measurement.
BERT:STAT ON
switches the BER measurement on.
Example:
TRIG:BERT
starts a single measurement.
Manual operation: See "State" on page 189
BERT and BLER Subsystems