Remote Control Commands
R&S
®
SMBV100A
508Operating Manual 1176.8016.02 ─ 17
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to SYSTem:PRESet.
Usage: Setting only
Manual operation: See "Preset" on page 118
*SAV <Number>
Save
Stores the current instrument settings under the specified number in an intermediate
memory. The settings can be recalled using the command *RCL with the associated
number.
To transfer the stored instrument settings in a file, use the com-
mand :MMEM:STOR:STAT.
Manual operation:
See " Save Immediate x" on page 162
*SRE <Contents>
Service request enable
Sets the service request enable register to the indicated value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<Contents> Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range: 0 to 255
*STB?
Status byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the TRIGger
subsystem.
Usage:
Event
Common Commands