Remote Control Basics
R&S
®
SMBV100A
468Operating Manual 1176.8016.02 ─ 17
6.2 Starting a Remote Control Session
The instrument and the controller have to be connected with the suitable cable and
switched on.
A remote control program must open a connection to the instrument (using VISA func-
tionality), before it can send commands to and receive device responses from the
instrument.
Instrument Address
In order to operate the instrument via remote control it must be addressed using the
defined interface address. See Chapter 6.1.3, "LAN Interface", on page 461, Chap-
ter 6.1.4, "USB Interface", on page 464, Chapter 6.1.5, "Serial Interface", on page 465,
Chapter 6.1.6, "GPIB Interface (IEC/IEEE Bus Interface)", on page 465 or Chap-
ter 6.1.7, "LXI Browser Interface", on page 467 for details.
The VISA resource strings are indicated in the "Setup > Remote Channel Settings"
menu.
Securing the display
To prevent unauthorized personnel from reading the display, you can disable the fre-
quency and level display explicitly. This is useful when you remotely control the instru-
ment from a different location.
For information on how to disable the frequency and level display, refer to "Security >
Annotation Frequency" and "Security > Annotation Amplitude" in the operating manual.
Refer to Chapter 6.2.3, "Examples", on page 470 for practical examples on setting up
of a remote control link and starting of a remote control session.
6.2.1 Switching to Remote Control
After switching on, the instrument is usually in the local state and can be operated via
the front panel controls (for instruments equipped with a display), a mouse and an
external keyboard.
Starting remote control
1. Send a command from a controller to the instrument.
The instrument changes to remote state as soon as it receives the command from
the controller.
Note: If you have sent &NREN before, the automatic transition from local state to
manual control by a subsequent remote command is disabled (use &GTR to enable
it again).
In remote state, operation via the front panel or via mouse and keyboard is dis-
abled. The status line indicates the "REMOTE" state.
Starting a Remote Control Session