5.5.8.1 NRP-Z Level Control Settings
To access the dialog for configuring the level control settings, perform one of the
following:
Select "RF/A Mod > config... > RF Level > NRP-Z Level Control".
Press the MENU key and select "RF/A Mod > RF Level > NRP-Z Level Con-
trol".
The dialog contains all parameters for configuring the settings for level control test
setup.
The remote commands required to define these settings are described in Chap-
ter 7.16.17, "SOURce:POWer Subsystem", on page 785
Sensor
Selects the R&S NRP-Z power sensor for power control.
Note: In remote control, the sensors are set up using the SENSe commands. The
remote measurement is triggered by the READ query which also provides the measure-
ment results.
The software version of the connected power sensor can be retrieved by means of the
remote control command :SENS:POW:TYPE?.
Use the "Setup >" Chapter 5.2.3.5, "NRP-Z Info/Update", on page 129 dialog to update
the sensor software.
Remote command:
[:SOURce<hw>]:POWer:SPC:SELect on page 793
State
Activates power control using the selected sensor.
The control loop periodically adjusts the generator output. After switching off, the run-
ning loop is completed.
Remote command:
[:SOURce<hw>]:POWer:SPC:STATe on page 794
Measured Level
Indicates the current reading of the sensor.
RF Block