Remote Control Commands
R&S
®
SMBV100A
543Operating Manual 1176.8016.02 ─ 17
:BLER:STOP
:BERT:STOP
Stops an ongoing measurement.
These commands automatically set BERT:STATe or BLER:STATe to OFF
.
Example:
BERT:STOP
stops the measurement. To start a new measurement, the BER
measurement must be switched on again by BERT:STATe ON.
Usage: Event
Manual operation: See "Terminated By" on page 190
:BLER:UNIT <Unit>
:BERT:UNIT <Unit>
Sets the unit for the error rate display on the screen. BERT|BLER:RES always speci-
fies the error rate as the quotient of "Number of errors" and "Number of data items",
unaffected by this command.
Parameters:
<Unit> ENGineering | SCIentific | PCT | PPM
ENGineering
The error rate is output in exponential notation with the exponent
–3 or –6.
SCIentific
The error rate is output in scientific notation, i.e. standardized to
one place to the left of the decimal.
PCT
The error rate is output in percent.
PPM
The error rate is output in parts per million.
*RST:
ENGineering
Example:
BERT:UNIT PPM
selects the unit "ppm" for the display of the error rate.
Manual operation: See "Bit Error Rate" on page 189
[:SOURce]:INPut:BERT:IMPedance <Impedance>
Sets the impedance of the BERT inputs CLOCK, DATA, DATA ENABLE and
RESTART.
Parameters:
<Impedance> G1K | G50 | G10K
*RST: G1K
BERT and BLER Subsystems