Remote Control Commands
R&S
®
SMBV100A
583Operating Manual 1176.8016.02 ─ 17
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
The command selects the filter length for user filter mode (SENSe:POWer:FILTer:
TYPE USER). As the filter length works as a multiplier for the time window, a constant
filter length results in a constant measurement time. Values 1 and 2^n are settable.
The time window is fixed to 20 ms.
Parameters:
<User> float
Range: 1 to 65536
*RST: 1
Example:
SENS:FILT:TYPE USER
selects user filter mode.
SENS:FILT:LENG 16
sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
Manual operation: See "Filter Length" on page 247
:SENSe<ch>[:POWer]:FILTer:NSRatio <NSRatio>
The command defines the noise content for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value determines the
proportion of intrinsic noise in the measured result.
Parameters:
<NSRatio> float
Range: 0.001 to 1
Increment: 0.001
*RST: 0.01
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR 0.2
sets a noise content of 0.2.
Manual operation: See "Noise Content" on page 247
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe <MTime>
The command defines the timeout for fixed noise filter mode
(:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio). This value ensures limited
settling times.
SENSe, READ and INITiate Subsystems