Use the "I/Q Mod In" section to select the input signal:
select "Analog Wideband I/Q In", if the external analog signal is fed into the I/Q
modulator.
The maximum available RF bandwidth is described in the instruments specifica-
tions (see data sheet).
select "Internal Baseband I/Q In", if the internally generated baseband signal is
used.
The RF bandwidth is reduced (see data sheet).
Alternatively, the input signal can be selected in the "I/Q Settings" menu (see Chap-
ter 5.6, "I/Q Modulation - I/Q Mod Block", on page 296).
5.6.2 I/Q Settings
To open the menu for setting the I/Q modulation parameters, select "I/Q Mod > Config-
ure > I/Q Settings" or use the MENU key.
System error correction of the I/Q modulator permits precise and repeatable measure-
ments. The correction routine should be called in the case of temperature fluctuations
of several degrees. To call the routine, press the SETUP key and select "System >
Internal Adjustment " and execute the desired adjustment procedure.
It is not possible to activate an I/Q modulation with an external analog signal ("Analog
Wideband I/Q In") simultaneously with:
AM,
BB-AM,
digital modulation,
digital standards,
noise,
arbitrary waveform generation and
multi carrier CW
I/Q modulation is activated and the source entered in the top section of the menu.
I/Q impairment for specific impairment of the I/Q modulation is set in the "Impairments"
section.
I/Q Modulation - I/Q Mod Block