Instrument Function
R&S
®
SMBV100A
204Operating Manual 1176.8016.02 ─ 17
Figure 5-7: Test generator cabling
1 = MARKER 1 connector
2 = MARKER 2 connector
3 = CLOCK OUT connector
4 = CLOCK IN connector
5 = NEXT connector
6 = TRIG connector
To establish the connections
1. Connect MARKER 1 and TRIG (1 -> 6) for data signal loopback.
2. Connect MARKER 2 and NEXT (2 -> 5) to loop back the data enable, or the restart
signal.
3. Connect CLOCK OUT and CLOCK IN (3 -> 4) to loop back the clock signal.
5.4.13.2 Test Generator Settings
Access:
► Select "Baseband > MISC > Bit Error Rate/Block Error Rate".
Bit and Block Error Rate Measurements