Remote Control Commands
R&S
®
SMBV100A
849Operating Manual 1176.8016.02 ─ 17
Manual operation: See " Execute Single Sweep - Frequency Sweep " on page 279
:TRIGger<hw>[:IMMediate]
The command immediately starts the activated sweep.
The command performs a single sweep and therefore applies to sweep mode AUTO
with sweep source SINGle. Use the commands
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING, and SOUR:SWE:FREQ|POW:MODE, or
SOUR:LFO:SWE:[FREQ:]MODE to set the respective sinlge sweep. You can alterna-
tively use an IMMediate command instead of the respective SWEep:
[FREQ:]|POW:EXECute command.
Example:
TRIG
starts all active sweeps.
Usage: Event
Manual operation: See " Execute Single Sweep " on page 266
7.21 UNIT Subsystem
The UNIT subsystem contains the commands specifying which units are valid if no unit
is indicated in a command. These settings are valid for the entire instrument.
:UNIT:ANGLe <Angle>
Defines the default unit for the phase modulation angle. It is not valid for commands
which determine angle values, e.g. RF phase. It does not influence the manual control
parameter unit and the display.
Parameters:
<Angle> DEGRee | RADian
*RST: RADian
Example:
UNIT:ANGL DEG
sets DEG as a default unit for all commands which determine
angle values.
:UNIT:POWer <Power>
Defines the default unit for power parameters. This setting affects the GUI, as well as
all remote control commands that determine power values.
Parameters:
<Power> V | DBUV | DBM
*RST: DBM
Example:
UNIT:POW V
sets V as a default unit for all commands which determine power
values.
UNIT Subsystem