Remote Control Commands
R&S
®
SMBV100A
671Operating Manual 1176.8016.02 ─ 17
Return values:
<MDelay> float
Range: 0 to 1
Increment: 1E-6
*RST: 0
Example:
BB:DM:MDEL?
queries the delay in the case of external modulation.
Response: 0.4 ms
the delay is 0.4 ms.
Usage: Query only
Manual operation: See "Modulation Delay" on page 359
Power Ramp
[:SOURce<hw>]:BB:DM:PRAMp:ATTenuation.................................................................. 671
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe]............................................................ 671
[:SOURce<hw>]:BB:DM:PRAMp:FDELay......................................................................... 672
[:SOURce<hw>]:BB:DM:PRAMp:RDELay.........................................................................672
[:SOURce<hw>]:BB:DM:PRAMp:SHAPe.......................................................................... 672
[:SOURce<hw>]:BB:DM:PRAMp:SOURce........................................................................673
[:SOURce<hw>]:BB:DM:PRAMp:TIME............................................................................. 673
[:SOURce<hw>]:BB:DM:PRAMp[:STATe].........................................................................673
[:SOURce<hw>]:BB:DM:PRAMp:ATTenuation <Attenuation>
Sets the level attenuation for signal ranges that are flagged with level attribute attenu-
ated by the LEV_ATT control signal.
Parameters:
<Attenuation> float
Range: 0 to 50
Increment: 0.1
*RST: 15
Default unit: dB
Example:
BB:DM:PRAM:ATT 15 dB
sets a level attenuation of 15 dB.
Manual operation: See "Attenuation" on page 371
[:SOURce<hw>]:BB:DM:PRAMp:BBONly[:STATe] <State>
This command selects power ramping in the baseband only or mixed power ramping in
the baseband and the RF section.
The ON setting is mandatory if, with power ramping active, only the baseband signal is
output (I/Q outputs)
Only then can a signal with a defined, predictable level be output.